Understanding The New Torque Wrench Calibration Standard

Torque Wrench Calibration Standard
Torque Wrench Calibration Standard
COMPARISON BETWEEN THE OLD (2003) AND NEW EDITION(2017) OF THE STANDARD ISO 6789

New torque wrench calibration standard published in two parts cancels and replaces ISO 6789:2003

ISO 6789-1:2017:

“Specifies requirements and methods for design conformance testing and quality conformance testing: minimum requirements for declaration of conformance”

ISO 6789-2:2017:

“Specifies the method for calibration of hand torque tools and determination of measurement uncertainty”

A. As per old standard ISO 6789:2003

[For type I: indicating torque tools and Type II: setting type torque tools]

A.1   No. of operation to be done and no. of readings  to be recorded for estimation of uncertainty and measurement deviation of a torque tool is as given below

Type & Class

Repeatability wre

Pre-Load

Readings at (20,60,100) %

Total

All type I

1

3 x 5

16

Type II classes A, B, D, E & G

5

3 x 5

20

Type II classes C & F

5

3 x 10

35

A.2 Rate of Loading

Minimum time period for application of torque values

For Type I:  Not specified in the standard

For Type II setting Torque Tools:

Force to be applied to the tool up to 80%. From 80% to the final target value, the force shall be applied slowly and uniformly during a period of 0.5 s to 4 s.

Torque wrench calibration standard

A.3 Components for determination of measurement uncertainty of the Torque Tool at 20%, 60% and 100% and measurement deviation

No. of computations

1. Repeatability (Type A)

wre

3

2. Uncertainty due to the variation in the scale, dial or display resolution

wr

3

3. Uncertainty of the measuring device

Wmd

3

4. Measurement deviation

As

3

Determination of the relative standard measurement uncertainty

W=

 

3

Total number of computations =

15

B. As per New Torque Wrench Calibration Standard ISO 6789-2:2017

 

B.1. No. of operations to be done and no. of readings to be recorded

as per the torque wrench calibration standard for estimation

of uncertainty of a torque tool is given below

 

 No. of readings to be taken to evaluate uncertainties due to
Repeatability wreReproducibility wrepOutput drive of the tool, wodInterface wintForce loading point wlTotal readings 
Pre-LoadReadingsReadingsPre-LoadReadingsPre-LoadReadingsPre-LoadReadings  
All type I, Type II classes A, D & G315(5 x 5)5(10 x 4)5(10 x 4)5(2 x 10)153 
Type II classes B, C, E, & F510(4 x 10)5(10 x 4)5(10 x 4)5(2 x 10)168 

B.2 Rate of Loading as per the New torque wrench calibration Standard ISO 6789-1:2017

Applied Torque value

<10 Nm

≥ 10 Nm

<100 Nm

≥100 Nm

<1000 Nm

≥1000 Nm

Minimum time to increase the torque from 80% of target value to the target value

0.5s

1s

1.5s

2s

For screw drivers (Type D, E and F) the time to increase the torque from 80% of target value to target value shall be between   

0.5s to 1.0s

For indicating tools Type I time intervals between any two subsequent loadings shall be of same duration

Setting torque tools (type II) Shall be loaded with a slowly and steadily increasing torque until attainment of the target torque is signaled by the torque tool. The target torque shall be reached after a minimum time is defined.

B.3 Components for determination of measurement uncertainty of the Torque Tool and Relative measurement error At tmin, 60% and 100%

No. of computations

Uncertainty due to the variation in the scale, dial or display resolution 

wr

3

Uncertainty due to reproducibility of torque tools

wrep

3

Uncertainty due to Geometric effects of the output drive of the tool

wod

3

Uncertainty due to Geometric effects of the interface between the output drive of the tool and the calibration system 

wint

3

Variation in Force loading point

wl

3

Repeatability (Type A)

wre

3

Uncertainty of the measuring device

Wmd

3

Relative measurement error at tmin, 60% & 100 %

1, 2, 3

3

Determination of the relative standard measurement uncertainty, W at tmin, 60% & 100 %

For Indicating Torque Tools: w+

3

For Setting Torque Tools: w+

Relative expanded uncertainty of measurement, W=k x w, where k = 2 coverage factor

 

Determination of the relative measurement uncertainty interval, W  

W  =  +W+   , Where,

  s is the mean value of relative measurement error and   =  100 in % [be is stated measurement error of the measurement device

3

Total number of computations =

30

Importance of avoiding parasitic forces

In the new torque wrench calibration standard this is the most significant change. The application of load is more clearly defined, and it is emphasized that the calibration device must allow the tool to move to prevent “parasitic” loads or moments. This means the connection between torque measurement system and torque tool shall permit self-alignment so that, the parasitic forces and moments are minimized.

Conclusion: Torque Wrench Calibration Standard

Torque wrench calibration standard as per new edition demands 153 to 168 operations, recording, controlling the rate of loading within specified time, 30 computations to arrive at uncertainty of measurements and the relative measurement error, and to ensure and to minimize parasitic loads or moments, it may be extremely difficult by manual operation [using manual torque calibrators] and computation of data manually.

Further, most of the present torque loaders (torque calibration devices) may not meet the above requirements and may become obsolete, hence may need to be computer-controlled or opt for a motorised torque calibration system for ease of use and computation.

We wish to inform you that Sushma Industries Pvt. Ltd. is already manufacturing motorized and or computerized Torque wrench calibration systems along with software to load and control the process of calibration, computation of results and preparation of calibration certificate as per ISO 6789-2:2017.

Sushma team will be pleased to assist you for consultancy, training, selection of new calibrators to comply with the new standard.

Need Help?

Lorem ipsum dolor sit amet, consectetur adipiscing elit. Ut elit tellus.

Latest Post

Concepts in Mass Metrology

Introduction There is a misconception that a weighing instrument measures weight. In fact, most accurate weighing instruments measure mass. Concepts in mass metrology states that

Read More »

Enquire Now